4 stdadd auto - measuring parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1104

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1088

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Selection

mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L

Default value

ppm

Control parameters

Dosing rate

Speed at which the volume increments are added.

Selection

slow | medium | fast

Default value

fast

Delta U

Potential difference that is to be achieved by a standard addition.

Range

1 ... 999 mV

Default value

10 mV

5.6.4.19.4

STDADD auto - Measuring parameters

Tab: Method

STDADD Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement with drift control
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a defined Waiting time has
elapsed.

Measured value drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s