Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 379

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Database

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

363

c2

Coefficient second-order of the Chebyshev polynomial.

c3

Coefficient third-order of the Chebyshev polynomial.

c4

Coefficient fourth-order of the Chebyshev polynomial.

User:

Name of the user logged in at the start of the determination.

Measuring input:

Shows the type of measuring input and serial number (e.g. 1.845.0010
11458 for iConnect).

Determination ID:

Unambiguous and unmistakable identification for the determination.

Colorimetric sensor
Information about the colorimetric sensor used.

Wavelength

Wavelength at which the calibration of the instrument has been carried
out.

c0

Calibration coefficient zeroth-order.

c1

Calibration coefficient first-order.

c2

Calibration coefficient second-order.

c3

Calibration coefficient third-order.

Confidence interval,

Measured values for the determination of the calibration curve must lie
within the confidence interval. Values outside this range are called outli-
ers, they are not integrated in the calculation of the calibration coefficient.