5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1080

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Device

Concentration unit of addition solution.

Selection

mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L

Default value

ppm

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

5.6.4.18.4

STDADD dos - Measuring parameters

Tab: Method

STDADD Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

Measurement with drift control
The measurement is stopped as soon as the Signal drift or the Stop
measured value
is reached or a defined Waiting time has elapsed.

Measured value drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for