5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

864

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.

Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.

Control parameters
Parameters for the control point.

Titration parameters
Parameters for the run of the titration.

Stop conditions
Definitions of conditions which cause the stop of the titration.

Monitoring
Definition of the monitoring of measured value, dosing rate and tem-
perature.

Evaluations
Definition of methods for evaluation of the titration curves.

Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated in the method run by the
command STAT and can be used in formulas under the designation
'Command name.Variable designation':

Designation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used
for the command

.DBL

Total duration for the processing of the command in s

.DSC

Time for processing all start conditions s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.ENP

Electrode zero point of the sensor used for the com-
mand (dimensionless)

.ETE

End temperature (temperature following processing of
the command) in °C