2 stdadd dos - general/hardware – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1091

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1075

Identifi-
cation

Description

.EME

Final measured value (measured value after processing of
the command) in the unit of the measured value

.ENP

Electrode zero point calculated from the calibration

.ETE

End temperature (temperature following processing of
the command) in °C

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

.IME

Initial measured value (measured value before start condi-
tions are processed) in the unit of the measured value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

.MTE

Temperature measurement with sensor; 1 = on, 0 = off)

.RES

Calculated result of the standard addition in the selected
unit

.SLO

Electrode slope calculated from the calibration

.STY

Type of stop with which the command was stopped: 1 =
normal
;0 = manual or after error

.TITER

Titer value of the solution used for the command

.VAR

Variance of the calculated result of the standard addition
in the selected unit

.VOL

Dosed volume

5.6.4.18.2

STDADD dos - General/Hardware

Tab: Method

STDADD Properties... General/Hardware

Command name

Name of the command.

Input

25 characters

The general parameters for the control device, the dosing device, the sen-
sor and the stirrer are defined on this tab.