5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

702

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Titrando (888)
Selection

1 | 2

Default value

1

855, 888
Selection

1

Default value

1

Titrino
Selection

1 | 2 | diff.

Default value

1

Sensor

Selection of a sensor of type pH electrode from the list of sensors availa-
ble in the sensor table. The calibration data for the sensor will be adopted
for the determination.

Selection

Sensor name | pH electrode

Default value

pH electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.