Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1652

Advertising
background image

7.20 IO controller

■■■■■■■■■■■■■■■■■■■■■■

1636

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Selection

'Port description' | Calculated value

'Port description'
All analog input ports can be selected.

0% Mapping

Measured value to be used when the Low level (0%) is applied as input
value.

Can be edited only for controlled variable = calculated value, other-
wise grayed-out and not available for editing.

Range

-214783647.000 ... 214783647.000

Default value

0 or controlled variable setting, respectively

100% Mapping

Measured value to be used when the High level (100%) is applied as
input value.

Can be edited only for controlled variable = calculated value, other-
wise grayed-out and not available for editing.

Range

-214783647.000 ... 214783647.000

Default value

100 or controlled variable setting, respec-
tively

Unit

Unit of the measured value to be used.

Can be edited only for controlled variable = calculated value, other-
wise grayed-out and not available for editing.

Input

50 characters

Default value

% or controlled variable unit, respectively

Default value

Selection or input of the standard value which is to be used for the input if
the result is invalid.

Range

0% mapping of the input ... 100% mapping
of the input

Selection

last value | setpoint value

Default value

last value

Setpoint value

Setpoint value to which the control is to be set.