5 det ipol - stop conditions, Stop conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 544

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

528

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

5.6.3.2.4.5

DET Ipol - Stop conditions

Tab: Method

DET Ipol Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the titration are defined on this tab. If more
than one stop condition is in effect at the same time, then the criterion
which is fulfilled first will stop the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Advertising