6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1410

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Note

Enter lower rates for high-viscosity liquids.

Tubing parameters
Defines the length and diameter of the tubing connected to the dosing
unit. The port assignment can also be altered. These parameters are
important for carrying out the dosing unit commands PREP and EMPTY
correctly, as the volumes of the tubing connections have to be taken into
account.

Note

Default values have already been entered for the tubing parameters;
these correspond to the dimensions of the standard tubing supplied. If
you do not make any alterations to the tubing connections then you do
not need to modify the tubing parameters. For information on length
and diameter of other tubings see Products/Accessories and col-
umns
under

http://products.metrohm.com

.

Dosing port 1

Port

Port to be used as Dosing port 1.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 1

Length

Length of tubing attached to Dosing port 1.

Range

0.0 ... 999.9 cm

Default value

40.0 cm

Diameter

Diameter of tubing attached to Dosing port 1.

Range

0.0 ... 9.9 mm

Default value

2.0 mm

Dosing port 2

Port

Port to be used as Dosing port 2.