Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1574

Advertising
background image

7.12 855 Robotic Titrosampler

■■■■■■■■■■■■■■■■■■■■■■

1558

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Min. beaker radius

Definition of the minimum radius which the beakers used on the rack
must have. If the lift is moved to the work position the beaker radii
defined in Rack table(see Chapter 6.12.2.1, page 1478) for general
Sample positions(see Chapter 6.12.3.2, page 1481) and Special beak-
ers
(see Chapter 6.12.3.4, page 1483) will be compared with the Min.
beaker radius
. If this minimum beaker radius is undercut then the run
will be stopped and an error message will appear. With off no check will
be made.

Selection

1.0 … 100.0 mm | off

Default value

off

Lift rate

Lift speed for manual operation.

Range

3 ... 25 mm/s

Default value

25 mm/s

Axial distance

Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.

Range

100.0 ... 300.0 mm

814 USB Sample Processor XL
Selection

166.0 mm

Default value

166.0 mm

815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler, 864
Balance Sample Processor
Selection

196.0 mm

Default value

196.0 mm

Swing Head
This parameter is only shown when a swing head is mounted on the
tower.

Serial number

Shows the serial number of the swing head.

Swing position

Lift position valid for all 4 external positions, at which the swing head
turns to the external positions.

Range

0 ... 235 mm

Default value

0 mm