5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

729

Max. rate

Rate at which dosing is carried out outside of the control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

1.00 (slow), 10.00 (optimal), 166.00
(fast) mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

1.00 (slow), 10.00 (optimal), 150.00
(fast) mL/min

Min. rate

Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. The smaller the
selected minimum rate, the slower the titration.

Range

0.01 ... 9999.00 µL/min

Default value

5.00 (slow), 25.00 (optimal), 50.00
(fast) µL/min

Stop criterion

Stop criterion

The titration us stopped after the last dosing once the endpoint has been
reached and the stop criterion has been fulfilled. The titration can be stop-
ped when a particular drift has been achieved or after a preset time.

Note

The Stop conditions will always stop the titration even if the stop crite-
rion has not been fulfilled or has been turned off.

Selection

drift | time | off

Default value

drift

drift
The switch-off point will always have the same curve slope dV/dt.
time
The endpoint must be exceeded for a certain time: the Delay time. The
same Delay time with different very small volume increments (depend-
ing on the volume of the exchange or dosing unit) means different
switch-off points.