5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1233

Selection

774 Oven Sample Processor | 874 Oven Sam-
ple Processor

Default value

874 Oven Sample Processor

Gas

Gas type

Gas type of the carrier gas.

Selection

Air | Nitrogen | other gas

Default value

Air

Gas flow factor

Factor for the correct determination of the gas flow. This parameter is
only visible for Gas type = other gas. The values for frequently used car-
rier gases are listed in the device manuals.

Range

0.001 ... 9.999

Default value

1.000

Gas flow

Inlet

Selection of the inlet of the carrier gas

Selection

Pump | Valve

Default value

Pump

Pump
Use of ambient air as carrier gas.
Valve
Use of gas out of a compressed gas cylinder.

Flow rate

874
Range

10 ... 150 mL/min

Default value

50 mL/min

Gas flow

Selection

Switch on | Switch off

Default value

Switch on

Switch on
Switches on the gas flow.
Switch off
Switches off the gas flow.