5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

560

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.5 ... 10.0 µA/mL

Default value

5.0 µA/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.5 ... 10.0 µA/mL

Default value

5.0 µA/mL

Break point evaluation

Note

This evaluation method is possible only with Titrando and 855.

Break-point evaluation

on | off (Default value: off)

A break-point evaluation is used to determine sharp changes of direction
in the titration curve.

EP criterion

Measure of the minimum sharpness of the break-point. The smaller the EP
criterion set, the more break-points will be found. As this is a relative value
related to the total measured value alteration, even small changes in the
measured value can be evaluated as a break-point for a small measured
value range.

Range

0 ... 1.0

Default value

0.3