Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1440

Advertising
background image

6.7 Subwindow Sensors

■■■■■■■■■■■■■■■■■■■■■■

1424

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Range

-2000.0 ... 2000.0 mV

Default value

0.0 mV

c (blank)

c (blank) is the third characteristic of the calibration function U = f(log
c)
. It reflects to a certain extent the curvature of the calibration function at
the lowest concentrations. This curvature is caused by the influence of so-
called interfering ions. If the concentration calibration is carried out with
less than 3 standards then c (blank) will be set equal to zero. The para-
meter c (blank) is available as a variable 'Command name.BLV' for cal-
culations.

Value
Range

0.00 ... 999999999

Default value

0.00

Unit
Input

10 characters

Default value

'empty'

Selection

mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L |
mg/mL | µg/mL

Specific calibration data for conductivity sensors

Cell constant

Cell constant of the conductivity sensor. This value can be determined
automatically with a calibration or entered manually. The cell constant is
available as a variable 'Command name.BLV' for calculations.

Range

0.001 ... 500 /cm

Default value

1 /cm

Specific calibration data for the Thermoprobe

a ... e

Display of the calibration coefficient a ... e.

Shared calibration data

Calibration temperature

Temperature during the calibration. The display also shows whether the
temperature was measured with a Pt1000 or an NTC temperature sensor
or entered manually (is not being displayed for the Thermoprobe).

Range

-20.0 ... 150.0 °C

Default value

25.0 °C