5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

631

Start measured value

Once the start measured value is reached, the dosing of the start volume
is stopped and the next start condition is processed or the titration is star-
ted. If the start measured value is achieved by the addition of a start vol-
ume, then titration will start directly.

Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.

Range

0.01 ... 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum

Start slope

Note

Only displayed for Titrando and 855.

Start slope

When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the dosing of a
start volume then the titration starts directly.

Range

0 ... 999 mV/mL

Selection

off

Default value

off

Dosing rate

Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.

Range

0.01 ... 166.00 mL/min

Default value

5.00 mL/min

Selection

maximum