3 properties - history – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1484

Advertising
background image

6.10 Subwindow Global variables

■■■■■■■■■■■■■■■■■■■■■■

1468

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted additionally to
the message defined above

Action
The actions can only be edited if monitoring for global variables is ena-
bled.

If during global variable monitoring it is found that the validity period has
expired then one of the following actions will be triggered automatically
at start test:

Selection

Document message | Display message | Cancel
determination

Default value

Display message

Document message
The message that the validity period of the global variable has expired
will be automatically saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period of the global variable has expired will be automati-
cally saved with the determination.
Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].

6.10.3.3

Properties - History

Tab: Configuration

Global variables [Edit] Properties Global variable -

'Name'

History

History table
The table contains the last 10 values for the selected global variable and
cannot be edited or sorted. The individual values are sorted according to
date so that the most up-to-date assignment is shown last.

Value

The values are shown in the following colors:

Blue, if the value has been assigned automatically by a method.

Black, when the values have been entered manually.

Orange, when the values are outside the warning limits.

Red, when the values are outside the intervention limits.

Advertising