5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

592

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

[Properties]

Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 931)
.

[Delete]

Delete the calculated measured value selected in the table.

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 932)
.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 932)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.3.2.9

MET pH - Measured value window

Dialog window: Method

MET pH Properties... Potentiometric evalua-

tion

Evaluation with measured value window (pH) [New]/[Proper-

ties]

Measured value window#

Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one endpoint will be recognized per window.

Lower limit pH

Lower limit for the measured value window.

Range

-20.000 ... 20.000

Default value

-20.000