Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1445

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1429

for ISE (+4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

17.0 mV

for ISE (-4)
Range

-2.147E+9 ... 2.147E+9 mV

Default value

-12.0 mV

pH(0) monitoring (for pH electrodes only)
on | off (Default value: off)

If this option is activated, then a check will be made to see whether the
electrode zero point pH(0) determined during the calibration of a pH elec-
trode is within the defined limits. If it is outside them, a corresponding
message appears and you can select whether you will nevertheless accept
the value or whether the previous value is to be retained.

Lower limit

Lower limit for pH(0).

Range

-2.147E+9 ... 2.147E+9

Default value

6.750

Upper limit

Upper limit for pH(0).

Range

-2.147E+9 ... 2.147E+9

Default value

7.250

E(0) monitoring (for ISE electrodes only)
on | off (Default value: off)

If this option is activated, then a check will be made to see whether the
electrode zero point E(0) determined during the calibration of an ISE elec-
trode is within the defined limits. If it is outside them, a corresponding
message appears and you can select whether you will nevertheless accept
the value or whether the previous value is to be retained.

Lower limit

Lower limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

-2000.0 mV

Upper limit

Upper limit for E(0).