5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1192

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Titrando, 855
Range

0.0 ... 2000.0 mV

Default value

2.0 mV

Upper limit

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

2000 mV

Upper hysteresis

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Range

0.0 ... 2000.0 mV

Default value

2.0 mV

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is execu-
ted.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a
message appears. As soon as the monitored measured value is again
within the limits (including hysteresis) reagent dosing can be resumed
by pressing the [Continue] button in this message box.