7 kft upol - conditioning, Conditioning – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 837

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

821

off
Stop when the entered time has elapsed after the start of the titration
(including start conditions).

Filling rate

Rate at which the dosing cylinder is to be filled after titration. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used.

Titrando
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.5.3.7

KFT Upol - Conditioning

Tab: Method

KFT Upol Properties... Conditioning

Command name

Name of the command.

Input

25 characters

Conditioning

Conditioning

on | off (Default value: off)

Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.

Start drift

The determination can only be started when the currently measured drift
is less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series
tab will then switch to COND READY.

Titrando
Range

1 ... 999 µg/min

Default value

20 µg/min