Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 34

Advertising
background image

1.11 What is new in tiamo 2.3 Patch 1?

■■■■■■■■■■■■■■■■■■■■■■

18

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

If a method was written in which the file name did not have the exten-
sion .pdf in the REPORT command at the field PDF file (completely
forgotten or only the period forgotten, e.g. abc or abcpdf), then this
would lead to a crash at the start of a determination.

If the digital output of the Avantes spectrometer was used (CTRL com-
mand), whereby a measuring command was also present in the
method that used the same device, then the message 014-122
Device occupied
would appear.

In the menu item View

Properties, the properties of the respective

live display have always been listed as Properties Live display
instead of Properties Live display #.

If it is determined during the Start Test that the wavelength of the cali-
bration of the colorimetric sensor does not match the wavelength of
the light source of the 089 Photometer in the command MEAS Opt
Conc
, then the message appears 014-042 Incorrect wavelength.
Instead of the value for the wavelength of the calibration of the colori-
metric sensor, the value of the wavelength of the light source of the
089 Photometer would be specified in the message.

The stop conditions for the commands DIGITAL IN and ANALOG IN
did not function, although the maximum running time did.

Program part Database

If a determination was carried out with intelligent electrodes, then the
ADC data (ADC type and serial number) of the analog (non-intelligent)
measuring input at which the iConnect was connected was stored in
the Database, subwindow Information on the tab Configuration.

If a determination was carried out with the 089 Photometer, then the
blank value was not subtracted in the database from the raw data IME
and EME (initial and end measured values, respectively) associated
with the commands MEAS Opt and MEAS Opt Conc.

If determinations that were linked with one another statistically that
also contained as results such things as the slope, the axis intercept
and the correlation coefficient for linear regressions (variables SLO,
ITC, COR) were recalculated in such a way that a determination or a
result was switched off, then the recalculated results of slope, axis
intercept and correlation were not correct.

Program part Method

An 867 pH module present in the device configuration could in fact be
selected as Device type in the command ELT MEAS, but not as
Device name. This led to a situation where the electrode test could
not be carried out with the 867 pH module.

The 846 Dosing Unit device could not be selected in the PORT com-
mand. This led to a situation where the PORT command could not be
performed with this kind of device, even though the firmware suppor-
ted this.

Advertising