Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 925

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

909

Monitoring Temperature

Monitoring Temperature

on | off (Default value: off)

If this check box is activated, then the temperature will be monitored and
any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the temperature. If the temperature falls below this limit,
the event Lower temperature limit exceeded is triggered.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

-20.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

-170.0 °C

Lower hysteresis

Lower hysteresis of the temperature. If the temperature once again
exceeds the lower limit by this hysteresis value, then the event Tempera-
ture lower limit ok
is triggered.

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Upper limit

Upper limit of the temperature. If the temperature exceeds this limit, the
event Temperature upper limit exceeded is triggered.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

150.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

500.0 °C

Upper hysteresis

Upper hysteresis of the temperature. If the temperature once again falls
below the upper limit by this hysteresis value, then the event Tempera-
ture upper limit ok
is triggered.