5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

562

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Window = Time
Range

0 ... 999999 s

Default value

999999 s

Gran evaluation

Note

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accord-
ance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 930).

Initial volume

Volume which is present in the measuring vessel before the command is
started.

Range

0.01 ... 9999.99 mL

Default value

50.00 mL

Lower limit

Value for the lower limit of the Gran evaluation range.

Range

-200.0 ... 200.0 µA

Default value

-200.0 µA

Upper limit

Value for the upper limit of the Gran evaluation range.

Range

-200.0 ... 200.0 µA

Default value

200.0 µA

5.6.3.2.5.8

DET Upol - Additional measured values

Tab: Method

DET Upol Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings: