5 met upol - stop conditions, Stop conditions – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 678

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

662

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

5.6.3.3.5.5

MET Upol - Stop conditions

Tab: Method

MET Upol Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop measured value

Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.

Range

-200.0 ... 200.0 µA

Selection

off

Stop EP

The titration is stopped when the specified number of equivalence points
has been found.

Range

1 ... 9

Default value

9

Selection

off

Advertising