2 curve display meas – Metrohm viva 1.0 (process analysis) User Manual

Page 287

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Database

viva 1.0 (for Process analysis)

■■■■■■■■

275

On

The check box in the Used column for the selected calibration point is activated.

Off

The check box in the Used column for the selected calibration point is deactiva-
ted.

All on

All check boxes in the Used column are activated.

All off

All check boxes in the Used column are deactivated.

Invert selection

The settings in the Used column for the selected calibration points are inverted.

Buttons

Save

The settings concerning the display of individual curves and the height
ratio between curve window and table are saved with this button.

Selection on

The marking of the curves selected in the table is activated with this but-
ton.

Selection off

The marking of the curves selected in the table is deactivated with this
button.

4.5.2.5.5.2.2

Curve display MEAS

Tabs
In the curve window a tab with the corresponding command name and
the run index (e.g. Chloride.1) is provided for each run of a measuring
command; their sequence is determined by the method.

Curve display for MEAS commands
The following elements are displayed in the upper part of the curve win-
dow:

Axes with labeling (measured quantity and unit)

Measurement curves

Zoom
Curves can be zoomed within the curve display as often as required by
spanning a square with the left mouse button pressed down. Zooming
can be undone by double-clicking or using the context-sensitive Show all
menu command.