3 cal meas opt - measuring parameters – Metrohm viva 1.0 (process analysis) User Manual

Page 482

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

470

■■■■■■■■

viva 1.0 (for Process analysis)

Displayed only when Device type = 089 Photometer.

Selection

continuous | off

Default value

continuous

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer
This section is displayed only when Device type = 089 Photometer.

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

3

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.5.2.6.4.3

CAL MEAS Opt - Measuring parameters

Tab: Method

CAL MEAS Opt Properties... CAL MEAS Opt - 'Command

name'

Measuring parameters

Command name

Name of the command.

Entry

25 characters

The measuring parameters for the calibration are defined on this tab.