8 Manual control

viva 1.0 (for Process analysis)

■■■■■■■■

903

Current Port

Shows the current port position.

Port

Selection of the target port to be switched to.

Input range

1 - 12

Default value

1

Rotating direction

Selection of the direction in which the valve moves to the target port.

Selection

ascending | descending | automatic | not over

Default value

ascending

ascending
The valve moves towards the target port in ascending direction.
descending
The valve moves towards the target port in descending direction.
automatic
The valve determines the shortest path to the target port autono-
mously.
not over
The valve moves towards the target port in such a way that the port
specified in the parameter Not over will not be traversed.

Not over

Selection of the port that is not to be traversed when the port is changed.
This parameter can be edited only if Rotating direction = not over.

Input range

1 - 12

Default value

1

[Start]

Sets the defined values.

Moves to current port position +1.

Moves to current port position –1.