5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

570

■■■■■■■■

viva 1.0 (for Process analysis)

Selection

'Controller name' | not defined

Default value

not defined

Controller type

Shows the type of the selected controller output.

Selection

Direct controller | Two-point controller | PID
controller
| PWM controller

Setpoint value

Setpoint value to which the control is to be set.

This parameter is not displayed when Controller type = Direct control-
ler
.

Input range

-214783647.000 - 214783647.000 'Mapping
unit of the analog input or of the calculated
value'

Default value

'empty' or default value from the selected
controller 'Mapping unit of the analog input
or of the calculated value'

Control variable

Shows the designation of the analog input port or of the calculated value
(formula entry).

This parameter is not displayed when Controller type = Direct control-
ler
.

Action
Selection of the controller action.

Selection

Switch on | Switch off | Duration

Default value

Switch on

Switch on
Activation of the regulation.
Switch off
Deactivation of the regulation.
Duration
Activation of the regulation during a defined period.

Duration

Time period during which controlling is to take place.

Input range

0.0 - 9,999.9

Default value

10.0

Selection

min | s

Default value

s