5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
376
■■■■■■■■
viva 1.0 (for Process analysis)
NOTE
No voltammetry commands (CVS) can be inserted in a series end track.
5.5.1.2.6
Exit track
Subwindow: Method
An exit track is a track that can be created manually by the user with
Insert
▶ New track... ▶ Exit track or the symbol. It starts with the
EXIT command (see Chapter 5.5.2.2.7, page 394) and ends with the
END command. Only one exit track can be inserted in a given method.
The exit track is executed in the run when all normal tracks and VA
tracks as well as the main track have been completed or when the
determination is manually canceled.