5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

525

814, 815, 858, 919
Selection

1 | 2 | 3

Default value

1

Set lines

Output signal

Entry of the bit pattern for the output signal or selection of a predefined
signal pattern.

It is possible to enter the following characters:

0 = Line inactive

1 = Line active

* = Any line status

p = Set pulse (pulse duration = 1 s). If a pulse with a different length is to
be output, then a corresponding template must be defined for this pur-
pose.

Sample changer, 894 and 846
Selection

Bit pattern with exactly 14 characters (0, 1, *,
p)
| ************** | Signal pattern

Default value

**************

Avantes Spectrometer
Selection

Bit pattern with exactly 10 characters (0, 1, *,
p)
| ********** | Signal pattern

Default value

**********

The bits are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Example: Control of the 843 Pump Station via Remote Box

Output signal

Function

****p*********

Activating pump 1 for 1 s

***p**********

Activating pump 2 for 1 s

****1*********

Activating pump 1

***1**********

Activating pump 2

****0*********

Deactivating pump 1

***0**********

Deactivating pump 2