5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

467

Confidence interval

Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers; they
are not integrated in the calculation of the calibration coefficient.

Selection

99 | 98 | 95 | off

Default value

98

off
The confidence interval will not be checked.

5.5.2.6.4

CAL MEAS Opt

5.5.2.6.4.1

CAL MEAS Opt - Overview

Dialog window: Method

CAL MEAS Opt Properties... CAL MEAS Opt - 'Com-

mand name'

Command for measuring standard solutions for the calibration of
colorimetric sensors
, used together with the CAL LOOP Opt com-
mand.

Instruments
This command can be executed with the following instruments:

Spectrometer: Avantes

Photometer: 089

Appearance
The command has the following appearance:

Parameters
The parameters for the CAL MEAS Opt command are set on the follow-
ing tabs:

General/Hardware
Selection of instrument type and colorimetric sensor.

Measuring parameters
Parameters for setting the measurement procedure.

Command variables
The following command variables are generated by the CAL MEAS Opt
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':