6 stir, 1 stir - overview, Stir – Metrohm viva 1.0 (process analysis) User Manual
Page 525
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0 (for Process analysis)
■■■■■■■■
513
2
Pump 2 is actuated.
1+2
The two pumps of the selected tower are actuated simultaneously.
Peristaltic
The peristaltic pump installed on the tower is actuated.
Rate
Speed at which the peristaltic pump is to be operated. This parameter is
visible only when Pump(s) = Peristaltic.
Input range
-15 - 15
Default value
10
Actions
Selection
Switch on | Switch off | Duration
Default value
Duration
Switch on
Activation of the pump(s).
Switch off
Deactivation of the pump(s).
Duration
Activation of the pumps during a defined period.
Pump(s) = 1, 2, 1+2
Input range
0 - 9,999.9
Default value
8.0
Pump(s) = Peristaltic
Input range
0 - 9,999.9
Default value
300.0
Selection
s | min
Default value
s
5.5.2.8.6
STIR
5.5.2.8.6.1
STIR - Overview
Dialog window: Method
▶ STIR ▶ Properties... ▶ STIR - 'Command name'
Command for the control of a connected stirrer.
Instruments
This command can be executed with the following instruments:
Dosing Interface: 846
Sample changer: 814, 815, 858, 919