3 meas ref - measuring parameters, Measuring parameters – Metrohm viva 1.0 (process analysis) User Manual

Page 438

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

426

■■■■■■■■

viva 1.0 (for Process analysis)

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

Avantes Spectrometer

Default value

Avantes Spectrometer

5.5.2.5.2.3

MEAS Ref - Measuring parameters

Tab: Method

MEAS Ref Properties... MEAS Ref - 'Command name' Mea-

suring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Start wavelength

Lower limit of the spectrum.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

End wavelength

Upper limit of the spectrum.

Input range

100.0 - 2,000.0 nm

Default value

1,000.0 nm

Integration time

Integration time for the acquisition of the spectrum.

Input range

0.01 - 600,000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are acquired and averaged.

Input range

1 - 10,000

Default value

10

Smoothing

Number of neighboring pixels with which the value for each measurement
pixel is determined.

Input range

0 - 100 pixels

Default value

0 pixels

Advertising