5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

520

■■■■■■■■

viva 1.0 (for Process analysis)

Device

Device name

Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

089 Photometer

Default value

089 Photometer

Heater

Target temperature

Temperature to which the cuvette has to be heated.

Input range

20 - 80 °C

Selection

Init. temp. | off

Default value

Init. temp.

Init. temp.
With Init. temp., the cuvette heater is activated at the same time as
the instrument and the cuvette is heated up to this temperature. The
initial temperature is entered and edited in the instrument configura-
tion (see Chapter 7.7.3, page 802).
off
Switch off heater.