Metrohm viva 1.0 (process analysis) User Manual

Page 485

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

473

Input range

0 - 999,999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula:

Input range

0 - 999,999 s

Default value

154 s

Measuring interval

Time interval for entering a measured value in the measuring point list.

Avantes Spectrometer
Input range

0.1 - 999,999.0 s (Increment: 0.1)

Default value

2.0 s

089 Photometer
Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Measuring time

Duration of the absorbance measurement.

Input range

0 - 999,999 s

Default value

120 s