5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

589

5.6.5

Evaluation - Calibration

5.6.5.1

Calibration- Overview

Subwindow section: Method

Evaluation - Calibration

Application
The way that the sample content for the substances is determined from
the peak data is defined in the Evaluation - Calibration section of the
evaluation. Only one calibration method can be selected for each method,
which is then valid for all available VA measuring commands.

Tabs

General

Calibration curves

5.6.5.2

Calibration - General

Tab: Method

Evaluation - Calibration General

This tab contains the basic settings for the calibration.

Calibration method

Selection of the method with which calibration is to take place.

Selection

no calibration | MLAT | DT

Default value

no calibration

MLAT
MLAT (Modified Linear Approximation Technique) is the default calibra-
tion method for the determination of brightener in electroplating
baths. Only the Linear regression curve type is permitted for this cali-
bration method (see Glossary, page 961).
DT
DT (Dilution Titration) is a calibration method for the determination of
suppressor in electroplating baths. The analysis is based on the fact
that galvanic metal precipitation is reduced by the addition of suppres-
sor to a VMS (Virgin Make-up Solution) (see Glossary, page 958).

5.6.5.3

Calibration - Calibration curves

Tab: Method

Evaluation - Calibration Calibration curves

Selection of the measuring command

Measuring command

Selection list with all of the VA measuring commands with data acquisi-
tion defined in the method.

Selection

'VA measuring command'