5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

493

imum speed depends on the cylinder volume of the dosing unit used. The
aspiration and expulsion of the sample should take place at a speed of <
10 mL/min.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.

5.5.2.7.7

PREP

5.5.2.7.7.1

PREP - Overview

Dialog window: Method

PREP Properties... PREP - 'Command name'

Command for rinsing the cylinder and tubings of a dosing unit.

Instruments
This command can be executed with the following instruments:

Voltammetry: 894

Dosing Interface: 846

Sample changer: 814, 815, 858, 919

Photometer: 089

Appearance
The command has the following appearance:

Parameters
The parameters for the PREP command are configured in the following
dialog window:

PREP - Properties

Command variables
The following command variables are generated by the PREP command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':