6 transfer, 1 transfer - overview, Transfer – Metrohm viva 1.0 (process analysis) User Manual

Page 547

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

535

NOTE

An event can only be received as such by the RECEIVE command if it
has been activated at the moment of the event, i.e. it must have been
started prior to the respective event. The occurrence of an event is sent
as a system message, only "listeners" who are already active will
receive, because it is not saved. A status, on the other hand, can be
requested at any time.

Comment

Comment on the event message

Entry

250 characters

Default value

'empty'

5.5.2.9.6

TRANSFER

5.5.2.9.6.1

TRANSFER - Overview

Dialog window: Method

TRANSFER Properties... TRANSFER - 'Command

name'

Command for data transfer via RS-232 to external instruments.

Appearance
The command has the following appearance:

Parameters
The parameters for the TRANSFER command are configured in the fol-
lowing dialog window:

TRANSFER - Properties

Command variables
The following command variables are generated by the TRANSFER com-
mand in the method run and can be used in formulas under the designa-
tion 'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started