5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

628

■■■■■■■■

viva 1.0 (for Process analysis)

5.6.7.4.5

Coefficient of determination R

2

Variable name

RS.’Command name’.’Substance name’.R2

Description

Coefficient of determination R

2

.

Decimal places

0…5…6

Unit

Calculation

Variable

Description

2

R

Coefficient of determination.

i

y

y value of calibration point i.

y

Mean value of all y values of the calibra-
tion points.

i

yˆ

The y value of calibration point i calcu-
lated with the calibration curve.

5.6.7.5

Volume results

5.6.7.5.1

Total volume

Variable name

RS.VTOT

Description

Current total volume in the measuring vessel.

Decimal places

0…3…6

Unit

#L

Remarks

The value of these variables is set to 0 at the start of a determination. The
respectively defined addition volume for the RS.VTOT variable is added
with each command ADD AUX (only if the option Include volume in
calculation
is activated), ADD STD, ADD SAMPLE and ADD SAMPLE