Metrohm viva 1.0 (process analysis) User Manual

Page 40

Advertising
background image

2.3 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

viva 1.0 (for Process analysis)

Designation

Description

Commands

OVS

Value of the output at the start of the command. Direct controller/2-point con-
troller command status: 1 = active, 0 = inactive

PID/PWM: -100 to 100% (value of the analog output or the pulse frequency)

CONTROL

PK{x}.ABS

Absorbance of the peak x (1 - 9) in mAU

MEAS Spec

PK{x}.TRN

Transmission of the peak x (1 - 9) in %

MEAS Spec

PK{x}.CNT

Intensity of the peak x (1 - 9) in counts

MEAS Spec

PK{x}.CNR

Intensity of the peak x (1 - 9) in the reference spectrum in counts

MEAS Spec

PK{x}.CND

Intensity of the peak x (1 - 9) in the dark spectrum in counts

MEAS Spec

PK{x}.WVL

Wavelength of the peak x (1 - 9) in nm

MEAS Spec

PK{x}.SAT

Detector with peak x (1 - 9) saturated: 1 = saturated;0 = not saturated

MEAS Spec

RPM

Stirring rate in min

-1

CVS

RAN

Current absolute rotation angle of the rack in ° in relation to the axis of the
selected tower (entry when exiting the command)

MOVE

RPO

Current rack position (entry when exiting the command); 0 means 'not defined'

MOVE

SAN

Current absolute swing angle of the robotic arm in ° (entry upon exiting the com-
mand)

MOVE, SWING

SDV

Actual value of the standard deviation when the command is canceled

CALL COND

SPO

Current external position (entry when ending the command); 0 means invalid
position

SWING

STC{x}

Concentration of the standard x (1 - 50) used for the command

CAL LOOP Opt

STY

Type of stop with which the command was stopped: 1 = normal; 0 = manual
or after error

CAL MEAS
Opt, CAL
Spec, CVS,
MEAS Opt,
MEAS Opt
Conc, MEAS
Ref, MEAS
Spec, MEAS
TMF

TMF

Transmission factor (dimensionless)

MEAS TMF

TOL

Tolerance within which a measured peak can be assigned to a calibration wave-
length.

CAL Spec

TOU

Timeout status: 1 = Max. waiting time expired; 0 = Max. waiting time not
expired

RECEIVE,
SCAN, TRANS-
FER

UNI

Mapping unit which is defined for an IO port in the configuration

ANALOG IN,
ANALOG OUT

Advertising