Metrohm viva 1.0 (process analysis) User Manual

Page 409

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

397

Devices
This command can be executed with the following devices:

Voltammetry: 894

Appearance
The command has the following appearance:

Principle
In the CVS (Cyclic Voltammetric Stripping) measuring mode, the potential
is changed once or more from the start potential to the vertex potential 1,
then to the vertex potential 2 and back to the start potential again at a
constant potential sweep rate. In this process, the current i is measured as
a function of the potential U.

t(i)

U

start

U

1

U

step

t

step

U

2

i

an

i

cath

U

U

t

U

step

Potential step (see Chap-
ter 5.5.2.3.2.4, page 407)

t

step

Potential step time (see
Chapter 5.5.2.3.2.4, page
407)

U

start

Start potential (see
Chapter 5.5.2.3.2.4, page
407)

U

1

1st vertex potential (see
Chapter 5.5.2.3.2.4, page
407)

U

2

2nd vertex potential
(see Chapter 5.5.2.3.2.4,
page 407)

i

an

Anodic current

i

cath

Cathodic current

Sequence of individual steps
In the CVS command, the following individual steps are successively exe-
cuted in the method run for the device defined in the command:

Advertising