Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

435

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate.

Input range

0 - 15

Default value

3

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.5.2.5.4.3

MEAS Opt - Measuring parameters

Tab: Method

MEAS Opt Properties... MEAS Opt - 'Command name' Mea-

suring parameters

Command name

Name of the command.

Entry

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measuring parameters for the Avantes spectrometer

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded.

Input range

100.0 - 2,000.0 nm

Default value

400.0 nm

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Input range

-1E+12 - 1E+12 (max. 10 digits) mAU

Default value

0 mAU

Integration time

Integration time for absorbance measurement.

Input range

0.01 - 600,000 ms

Default value

6 ms