Metrohm viva 1.0 (process analysis) User Manual
Page 570
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
558
■■■■■■■■
viva 1.0 (for Process analysis)
Start delay
Waiting time in seconds or minutes until the triggered action is started.
Input range
0 - 214783647
Default value
0
Selection
s | min
Default value
s
Working time
The time for which the digital output is set to active can be specified in
seconds or minutes for Action = Timer. After the working time has
elapsed, it is set to inactive again.
Input range
0.01 - 999,999
Default value
0
Selection
s | min
Default value
s
End delay
Waiting time after the working time has expired.
Input range
0 - 214783647
Default value
0
Selection
s | min
Default value
s
Parameters
Parameter 1
The meaning of Parameter 1 depends on the Type set for the digital
output.
For Type = Stirrer: stirring performance in %, pulse duration dependent
on Parameter 2.
Input range
-214,783,647 - 214,783,647
Default value
100
Parameter 2
The meaning of Parameter 2 depends on the Type set for the digital
output.
For Type = Stirrer:
■
0: Duration of the switch-on pulse set to 500 ms.