5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

558

■■■■■■■■

viva 1.0 (for Process analysis)

Start delay

Waiting time in seconds or minutes until the triggered action is started.

Input range

0 - 214783647

Default value

0

Selection

s | min

Default value

s

Working time

The time for which the digital output is set to active can be specified in
seconds or minutes for Action = Timer. After the working time has
elapsed, it is set to inactive again.

Input range

0.01 - 999,999

Default value

0

Selection

s | min

Default value

s

End delay

Waiting time after the working time has expired.

Input range

0 - 214783647

Default value

0

Selection

s | min

Default value

s

Parameters

Parameter 1

The meaning of Parameter 1 depends on the Type set for the digital
output.

For Type = Stirrer: stirring performance in %, pulse duration dependent
on Parameter 2.

Input range

-214,783,647 - 214,783,647

Default value

100

Parameter 2

The meaning of Parameter 2 depends on the Type set for the digital
output.

For Type = Stirrer:

0: Duration of the switch-on pulse set to 500 ms.