5.6 Evaluation subwindow

■■■■■■■■■■■■■■■■■■■■■■

582

■■■■■■■■

viva 1.0 (for Process analysis)

Tolerance

Tolerance of the verification potential for the assignment of a peak to a
substance. The peak potential determined must be within the range Char-
acteristic potential
± Tolerance so that the substance is recognized.

Input range

0.001 - 1.000 V

Default value

0.200 V

Min. width

Minimum peak width for the assignment of a peak to a substance. The
calculated peak width must be larger than Min. width in order for the
substance to be recognized.

Input range

0.001 - (Max. width) V

Default value

0.010 V

Max. width

Maximum peak width for the assignment of a peak to a substance. The
calculated peak width must be smaller than Max. width in order for the
substance to be recognized.

Input range

(Min. width) - 1.000 V

Default value

0.500 V

Min. measured quantity

Minimum measured quantity for the assignment of a peak to a substance.
The measured quantity must be greater than Min. measured quantity
in order for the substance to be recognized. Value and unit can be
entered or selected separately.

Input range

1 - 999

Default value

200

Selection

mA | µA | nA | pA

Default value

pA

Sweep direction

Specification as to whether the substance is to be evaluated during for-
ward or reverse movement of the cyclical sweep.

Selection

anodic | cathodic

Default value

anodic

Navigation bar

Navigation bar for changing the displayed substances with the following
functions: