Metrohm viva 1.0 (process analysis) User Manual

Page 420

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

408

■■■■■■■■

viva 1.0 (for Process analysis)

1st vertex potential

First inflection point for the sweep.

Condition:1st vertex potential

Start potential

Input range

-5.000 - 5.000 V (Increment: 0.001)

Default value

-0.175 V

2nd vertex potential

Second inflection point for the sweep.

Condition 1: 2nd vertex potential

1st vertex potential

Condition 2 (for anodic sweeps): 1st vertex potential > Start poten-
tial

2nd vertex potential

Condition 2 (for cathodic sweeps): 1st vertex potential < Start poten-
tial

2nd vertex potential

Input range

-5.000 - 5.000 V (Increment: 0.001)

Default value

1.625 V

Potential step

Potential step for the DC potential ramp.

Condition 1: Potential step

≤ (|2nd vertex potential1st vertex

potential

|) / n

min

n

min

(minimum number of measuring points for the evaluation) depends

on the Potential step itself as follows:

Potential step

≥ 0.0095 V: n

min

= 16

Potential step < 0.0095 V and

≥ 0.0075 V: n

min

= 20

Potential step < 0.0075 V and

≥ 0.0055 V: n

min

= 24

Potential step < 0.0055 V and

≥ 0.0035 V: n

min

= 32

Potential step < 0.035 V : n

min

= 52

Condition 2: Potential step

≥ (|2nd vertex potential1st vertex

potential

|) / 65,535

Input range

0.00015 - 5.00000 V (Increment: 0.00001)

Default value

0.006 V

Potential step time

Shows the time that has to expire before the potential is changed by the
amount of Potential step.

Calculation: Potential step / Sweep rate

Sweep rate

Potential sweep rate for the sweep.