Metrohm viva 1.0 (process analysis) User Manual

Page 736

Advertising
background image

6.8 Colorimetric sensors subwindow

■■■■■■■■■■■■■■■■■■■■■■

724

■■■■■■■■

viva 1.0 (for Process analysis)

breach of these limits does not trigger any action. The limits set on the
Limit values tab apply for the monitoring during the calibration.

Warning limits for coefficient of determination
on | off (Default value: off)

If this option is enabled, then the values for the coefficient of determina-
tion are displayed in orange if these limits are breached.

Lower limit

Lower warning limit for the coefficient of determination.

Input range

0.00000 - 1.00000

Default value

0.98000

Upper limit

Upper warning limit for the coefficient of determination.

Input range

0.00000 - 1.00000

Default value

1.00000

Intervention limits for coefficient of determination
on | off (Default value: off)

If this option is enabled, then the values for the coefficient of determina-
tion are displayed in red if these limits are breached.

Lower limit

Lower intervention limit for the coefficient of determination.

Input range

0.00000 - 1.00000

Default value

0.96000

Upper limit

Upper intervention limit for the coefficient of determination.

Input range

0.00000 - 1.00000

Default value

1.00000

Advertising