5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

409

Condition 1: Sweep rate

Potential step / 10 s

Condition 2: Sweep rate

Potential step / 0.001 s

Input range

0.001 - 10.000 V/s (Increment: 0.001)

Default value

0.100 V/s

Preparation cycles

Number of cycles that are run through before the measuring cycles with-
out current measurement.

Input range

0 - 200 (Increment: 1)

Default value

1

Measuring cycles

Number of cycles with current measurement.

Input range

1 - 50

Default value

2

Sweep duration

Shows the duration of the sweep in s.

Calculation: (Preparation cycles + Measuring cycles)

Ч [ 2 Ч (| 2nd

vertex potential

1st vertex potential |) / Potential step ] × Poten-

tial step time

NOTE

The actual measurement of potential and current takes place at the end
of the Potential step time. The measuring time is defined as follows:

Potential step time < 0.080 s => measuring time = Potential
step time
/4 s

Potential step time > 0.080 s => measuring time = 0.02 s (at a
line frequency of 50 Hz)

Potential step time > 0.080 s => measuring time = 0.01667 s (at a
line frequency of 60 Hz)