5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

484

■■■■■■■■

viva 1.0 (for Process analysis)

not defined
The dosing unit must be assigned with not defined by the user at the
start of the method.

Dosing rate

Rate (volume per time unit) of expulsion. The maximum rate depends on
the cylinder volume of the buret used and on the diameter of the hose at
the selected dosing port.

Input range

0.01 - 166.00 mL/min

Default value

2.00 mL/min

Selection

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.

Filling rate

Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the buret used and on the diameter of the hose at the
selected fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

5.5.2.7.5

ADD SAMPLE DT

5.5.2.7.5.1

ADD SAMPLE DT - Overview

Dialog window: Method

ADD SAMPLE DT Properties... ADD SAMPLE DT -

'Command name'

Command for the addition of small sample volumes for the calibra-
tion method DT (Dilution Titration) for the determination of suppressor in
electroplating baths.

Appearance
The command has the following appearance:

Parameters
The parameters for the ADD SAMPLE DT command are configured in
the following dialog window: