5 Method

viva 1.0 (for Process analysis)

■■■■■■■■

387

NOTE

The sample data variables Analytical volume and Dilution volume
must both be either present (sample dilution) or absent (no sample dilu-
tion) as variables in the START command.

Fixed value

on | off (Default value: off)

If this check box is activated, the sample data variable can be assigned a
fixed value. For variables of the Date/Time type, the date has to be
entered in the 2.4.1Selecting the date dialog window. No formulas can
be entered in this field.

Type = Number
Input range

-1.0E+99 - 1.0E+99

Type = Text
Entry

100 characters

Type = Date/Time
Selection

'Date'

NOTE

Variables of the Text type with activated fixed value assignment can be
used for saving data received with a TRANSFER command in these
variables (see Chapter 5.5.2.9.6.3, page 537).

Check at start

on | off (Default value: on)

If this check box is activated, whether the sample data variable is valid and
whether the limit values are maintained with activated monitoring is
checked during the start test.

NOTE

Check at start always has to be activated for the sample data varia-
bles Sample amount, Analytical volume and Dilution volume in
order for a determination to run.