5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

510

■■■■■■■■

viva 1.0 (for Process analysis)

Selection

814 USB Sample Processor | 815 Robotic USB
Sample Processor XL
| 858.0010 Professional
Sample Processor
| 858.0020 Professional
Sample Processor
| 858.0030 Professional
Sample Processor
| 919.0020 IC Autosampler
plus

Default value

858.0020 Professional Sample Processor

Target

Tower

Selection of the tower on the sample changer for moving to the required
lift position.

Selection

1 | 2

Default value

1

Lift position

Entering the desired lift position in mm or selection of one of the lift posi-
tions predefined for the rack. A lift height of 0 mm corresponds to Home
position
. Here the lift is moved to the uppermost stop.

Input range

0 - 235 mm

Selection

Work position | Shift position | Rinse position |
Special position | Home position

Default value

Work position

NOTE

Please take care that no lift positions exceed the maximum stroke path
specified in the tower properties of the instrument.

Parameters

Lift rate

Selection of the speed for moving the lift.

Input range

5 - 25 mm/s

Default value

25 mm/s